Shanghai Changai Electronic Science & Technology Co., Ltd.
Products
Contact Us
  • Contact Person : Mr. Gao Ben
  • Company Name : Shanghai Changai Electronic Science & Technology Co., Ltd.
  • Tel : 86-21-51068326
  • Address : Shanghai,Shanghai,13D, 1138 Changan Road, Zhabei
  • Country/Region : China

GNL-YF1 Oxygen Analyzer

GNL-YF1 Oxygen Analyzer
Product Detailed
Related Categories:Gas Analyzers
* Specially used in wave soldering & reflow soldering oven to monitor the oxygen concentraton * 4-20mA or 0-5V DC signal output

1. Introduction                                                                                                 

New product GNL-YF1 analyzer (only for wave soldering oven and reflow soldering oven) breaks through traditional magneto mechanical measurement mode and measures oxygen concentration without benchmark gas by adopting conversion ion current with the world advanced technology as measurement principle. It has observable characteristics of long calibrating period, high precision, small appearance and low power consumption.

2. Features                                                                                                   

* New pattern, long-life, conversion ion current tester with fast response and high sensitivity

* Concise menu, simple operation

* Checking the sensor working state automatically and unique double-line working change mode extending the sensors life

* Setting upper and lower alarm values freely during the measure ranges

* 4-20mA or 0-5V DC signal outputs

* It can achieve two-way communication by connected to a computer via the standard RS232C or RS485 interface.  

3. Technical Specifications                                                                                      

Measure Range:                   0-10ppm, 0-100ppm, 0-1000ppm, 0-25% O2; auto-ranging

Accuracy:                           ±0.5ppm in the range of 0-10pppm;

                                          ±1ppm in the range of 0-100pppm;

                                          ±2ppm in the range of 0-1000pppm;

                                          ±1% in the range of 0-25%

Stability:                             Zero drift ≤±1%FS/7d

                                          Span drift ≤±1%FS/7d

Contact Capacity:                125VAC 0.3A, 110VDC 0.3A, 30VDC 1A

Repeatability:                      ≤±1%

Response Time:                  90% of FS reading ≤10 seconds

Sampling System:               Film sampling pump or inner minus pressure system

Sample Gas Flow:               200300ml/mim

Ambient Temp.:                   -10oC~+50oC

Power Supply:                     85V-265V AC, 50Hz±5%

Sensor Life:                        36 months at 25 and 1 atm

Warranty:                          12 months analyzer; 12 months sensor

Weight:                              3kg

Installation Dimension:         125mm×200mm×300mm

4. Applications                                                                                               

It can be used to measure the oxygen concentration in the fields of magnetic materials, high temperature sintering furnace, the electronics industry protective gas.

5. Order knowledge                                                                                           

Measure range

Tested gas pressure

Gas composition and impurity

GNL-YF1 Oxygen Analyzer



Copyright Notice @ 2008-2022 ECBAY Limited and/or its subsidiaries and licensors. All rights reserved.